CPU board failures cause fowler (backrest) electronic controls to stop working.
Source facts
Action
Stryker Medical initiated a voluntary recall of the S3 MedSurg Bed, Model 3005, with notices sent on February 2, 2016 via certified mail. Stryker medical identified that Fowler does not raise up to desired angle; and the Fowler does not raise electronically, but manual function remains operational through the manual CPR release. Actions Needed for Customers:
1. Locate the bed(s) listed by serial number on the attached business reply form.
2. A Stryker ProCare Representative will contact your facility to schedule a time for repair.
3. Fax (269)488-8691 or email productfieldaction@stryker.com the enclosed business reply form to confirm receipt of this notification.
4. If you have further distributed the beds listed in this letter, please forward a copy of this notice to the new users and advise us of their new location. If you have any urgent questions or concerns, please, contact Stryker Customer Service at (1-800-327-0770). Normal business hours are Monday-Friday 8 a.m-5 p.m. (ET).
Initiated
2016-02-02
Root cause
Component change control
Terminated
2017-06-12
Distribution
Distribution US nationwide, Canada, and Mexico.
Recall number
Z-2353-2016
Recall status
Terminated
CDRH record ID
143629
FDA classified
2016-08-08
Recall event number
73255