Skip to content

Public FDA record

Z-2353-2024: QIAcube Connect MDx - IVD Instrument designed to perform automated isolation and purification of nucleic

Recall record Z-2353-2024 for Qiagen Sciences LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2353-2024
Company in FDA record
Qiagen Sciences LLC
FDA classification date
Jul 12, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Heater shaker module does not perform heating if the temperature is set below 40C for static incubation steps when running IVD protocol QIAamp DSP DNA Mini Kit (cat. no. 61304), Protocol: Isolation of genomic DNA from Gram-positive bacteria (Script filename: Kt_name_Bacteria (Gram+) or rY'east_Enzymatic Lysis_ V2) which could potentially cause delayed or erroneous results depending on the downstream assay.

Source facts

Action
Qiagen issued Urgent Medical Device Correction Letter to Lab Director/Manager, Medical Director, Risk Manager, Safety Officer via email on June 3, 2024. Letter states reason for recall, health risk and action to take: l Complete the Acknowledgment of Receipt promptly and return to quality.communications@qiagen.com l If running the above protocols and you have selected an incubation temperature for the bacterial lysis below 40C, then you can ensure precise temperature control by using the heater shaker built into the QIAcube Connect MDx via the Tools menu of the software. If you wish to check the actual applied temperature in your workflow as currently used, we offer to analyze your logfiles to determine the actual applied temperature. This may help you to confirm these settings. Please contact QIAGEN Technical Services at (800 362 7737) for assistance,
Initiated
2024-06-03
Root cause
Software design
Distribution
Nationwide. Foreign: AD, AT, AU, BE, CA, CH, CN, CZ, DE, DK, ES, FI, FR, GB, ID, IN, IT, JO, KR, KZ, LU, ML, MY, NL, NO, PH, PL, PY, SA, SE, SG, SK, SO, VN J, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI
Recall number
Z-2353-2024
Recall status
Open, Classified
CDRH record ID
208453
FDA classified
2024-07-12
Recall event number
94836

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗