Open original FDA record ↗Recall record
- FDA identifier
- Z-2354-2020
- Company in FDA record
- Acuity Surgical Devices, LLC
- FDA record creation date
- Jun 11, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to a manufacturing process issue, it allowed vertebral body bone screws to not seat properly within the spinal device/cage.
Source facts
- Action
- On 12/14/2018, the firm notified its supplier via email requesting to visually review its inventory which determined that some screw holes were oval shaped. On 7/23/2019, all inventory was quarantined as part of the removal and was scrapped from inventory. Due to the low risk to patients and low probability, the firm ownly initially begin removal of product from internal locations (main warehouse) for the purpose of inspection and if necessary, repair or destruction. Removal Summary No additional customer or distributor communication took place during the removal.
- Initiated
- 2019-01-07
- Root cause
- Process control
- Distribution
- US Nationwide distribution including in the states of TX, FL, NC, LA, KY, WI, SD, IL, CA, and KS. O.U.S.: None
- Recall number
- Z-2354-2020
- Recall status
- Open, Classified
- CDRH record ID
- 181452
- FDA record created
- 2020-06-11
- Recall event number
- 85642