Skip to content

Public FDA record

Z-2355-2020: A LINK Z INTERVERTEBRAL BODY FUSION DEVICE Product Description: A Link Z Ti Unitary Open Interbody 32x24

Recall record Z-2355-2020 for Acuity Surgical Devices, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2355-2020
Company in FDA record
Acuity Surgical Devices, LLC
FDA record creation date
Jun 11, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to a manufacturing process issue, it allowed vertebral body bone screws to not seat properly within the spinal device/cage.

Source facts

Action
On 12/14/2018, the firm notified its supplier via email requesting to visually review its inventory which determined that some screw holes were oval shaped. On 7/23/2019, all inventory was quarantined as part of the removal and was scrapped from inventory. Due to the low risk to patients and low probability, the firm ownly initially begin removal of product from internal locations (main warehouse) for the purpose of inspection and if necessary, repair or destruction. Removal Summary No additional customer or distributor communication took place during the removal.
Initiated
2019-01-07
Root cause
Process control
Distribution
US Nationwide distribution including in the states of TX, FL, NC, LA, KY, WI, SD, IL, CA, and KS. O.U.S.: None
Recall number
Z-2355-2020
Recall status
Open, Classified
CDRH record ID
181480
FDA record created
2020-06-11
Recall event number
85642

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.