Cincinnati Sub-Zero Products LLC, a Gentherm Company
FDA record creation date
Sep 4, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
The manual does not provide instruction stating that exceeding 40 Celsius for extended periods can cause tissue damage.
Source facts
Action
1. Access updated manuals and ensure that obsolete manuals are removed from services. updated manuals may be accessed via www.gentherm.com or physical copies may be requested from Gentherm Medical, LLC at 1-888-437-5608
2. Inform the content of the notification to all users and if you have further distributed this product, please provide those account with a copy of this notice.
3. Please complete and return the enclosed response form as soon as receipt of this notification and to inform Gentherm Medical, LLC and return the form by e-mail to FA2019-005@gentherm.com or mail to : Gentherm Medical, LLC, 12011 Mosteller Road, Cincinnati, OH 45241.
If you have any questions, call Gentherm Medical Technical Support at 1-888-437-5608.
Initiated
2019-06-20
Root cause
Labeling False and Misleading
Terminated
2020-06-09
Distribution
Ship-To State
AL
AR
AZ
CA
CO
CT
DC
DE
FL
GA
HI
IL
IN
KS
KY
LA
MA
MD
ME
MI
MN
MO
NC
ND
NE
NH
NJ
NM
NY
OH
OK
PA
RI
SC
TN
TW
TX
UT
VA
WA
WI
WV
WY
Ship-To Country
United States
Taiwan
Egypt
United Arab Emirates
Israel
Pakistan
Turkey
Colombia
Thailand
Kuwait
United Kingdom
Mexico
Netherlands
Canada
Ecuador
South Korea
Germany
Georgia
China
Panama
Japan
Malaysia
Hong Kong
Austria
Chile
Mongolia
Morocco
Australia
Indonesia
Venezuela
Singapore
Recall number
Z-2356-2019
Recall status
Terminated
CDRH record ID
174995
FDA record created
2019-09-04
Recall event number
83376
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.