Skip to content

Public FDA record

Z-2357-2008: GE Innova 2100IQ Versatile Cardiac and Vascular Imaging System, GE Healthcare, Waukesha, WI 53188 ( The s

Recall record Z-2357-2008 for GE Healthcare. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2357-2008
Company in FDA record
GE Healthcare
FDA classification date
Sep 21, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

GE Healthcare has become aware of 2 potential conditions that could occur with the MAVIG GD Monitor Suspension that is used with Advantx-E, Innova 2000, 21000IQ, 3100/3100IQ, 4100/4100IQ and 2121 IQ/3131IQ cardiovascular systems. The two conditions include 1). Due to insufficient securing of the connecting elements, the LCD vertical monitor support may disengage from its arm and fall on the tab

Source facts

Action
The consignees were sent a "GE Urgent Medical Device Correction" letter dated May 2008. The letter was addressed to Manager of Radiology/Cardiology; Radiologists/Cardiologists and; Biomedical/Clinical Engineering. The letter described the problem, affected products, short term user recommendations, GE Healthcare resolution plan and contact information. Contact GE Healthcare at 1-262-544-3894 for assistance.
Initiated
2008-06-24
Root cause
Employee error
Terminated
2011-12-17
Distribution
USA: AR, CA, GA, IN, LA, MA, MD, MO, NJ, NM, NV, NY, OH, OR, PA, SC, TN TX, WI, AL, AZ, CO, CT, DE, FL, HI, IA, ID, IL, IN, KS, KY, LA, ME, MI, MN, MS, MT, NC, ND, NE, NH, OK, OR, PA, SD, UT, VA, VT, WA,WV, KY, and WI OUS: ALGERIA AUSTRALIA AUSTRIA BELARUS BELGIUM BRASIL CANADA CHILE CHINA CHINA COLUMBIA CZECH REPUBLIC DENMARK DV EGYPT FINLAND FRANCE GEORGIA GERMANY GREECE GUATAMALA HONG KONG HUNGARY INDIA INDONESIA IRAN IRAQ IRELAND ISRAEL ITALY JAPAN KOREA KUWAIT LATVIA LEBANON MACEDONIA MALAYSIA MEXICO MOROCCO NETHERLANDS NEW ZEALAND NORWAY PAKISTAN PANAMA POLAND PORTUGAL PUERTO RICO RUSSIA SAUDI ARABIA SINGAPORE SLOVAKIA SPAIN SUDAN SWEDEN SWITZERLAND SYRIA TAIWAN TUNISIA TURKEY UNITED ARAB EMIRATES UNITED KINGDOM and VENEZUELA
Recall number
Z-2357-2008
Recall status
Terminated
CDRH record ID
72408
FDA classified
2008-09-21
Recall event number
48961

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗