The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.
Source facts
Action
On July 26, 2018, the firm sent an email to it's US distributor that requested a sub-recall.
Initiated
2018-07-26
Root cause
Nonconforming Material/Component
Distribution
Illinois
Recall number
Z-2357-2019
Recall status
Open, Classified
CDRH record ID
172946
FDA record created
2019-08-22
Recall event number
82867