Z-2357-2019: Microdot Xtra blood glucose test strips

Recall record Z-2357-2019 for Cambridge Sensors Limited. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2357-2019
Company in FDA record
Cambridge Sensors Limited
FDA record creation date
Aug 22, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.

Source facts

Action
On July 26, 2018, the firm sent an email to it's US distributor that requested a sub-recall.
Initiated
2018-07-26
Root cause
Nonconforming Material/Component
Distribution
Illinois
Recall number
Z-2357-2019
Recall status
Open, Classified
CDRH record ID
172946
FDA record created
2019-08-22
Recall event number
82867

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗