Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or under exposed image.
Source facts
Action
Philips sent customers an Electronic Product Radiation Correction letter dated July 14, 2021. The letter identified the affected product, problem and actions to be taken.
Philips plans to correct the defect at no cost to customers by implementing a field correction involving the installation of a software update. For questions contact the Customer Care Solutions Center 1-800-722-9377 and reference FCO70900051.
Initiated
2020-07-22
Root cause
Radiation Control for Health and Safety Act
Distribution
US Nationwide Distribution
Recall number
Z-2357-2021
Recall status
Open, Classified
CDRH record ID
188563
FDA record created
2021-08-28
Recall event number
88393