Z-2357-2024: Brand Name: Meridian Angled Awl Fixed Product Name: Angled Awl Fixed Model/Catalog Number: RA2-003015 Pro

Recall record Z-2357-2024 for SEASPINE ORTHOPEDICS CORPORATION. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2357-2024
Company in FDA record
SEASPINE ORTHOPEDICS CORPORATION
FDA classification date
Jul 15, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Insufficient weld around the cap component of the instrument is insufficient to withstand impaction forces.

Source facts

Action
On June 4, 2024, SeaSpine Orthopedics issued a "Urgent Medical Device Recall" Notification via E-Mail to affected consignees. SeaSpine Orthopedics ask consignees to take the following actions: We request that you return the affected part to SeaSpine s Carlsbad facility for evaluation and potential rework. To support your upcoming cases, a replacement Vu aPod Prime NM Angled Awl will be sent (PN: 33-40-1028). Additionally, please continue to utilize the loaner pool. Please use the attached label to return the inventory in your possession at your earliest convenience. Once the inventory has been returned, please acknowledge via DocuSign.
Initiated
2024-06-04
Root cause
Process design
Distribution
The product was distributed to the following US states: Nevada, Colorado, Texas, Washington, Indiana, Georgia, Michigan, Florida, Pennsylvania, Illinois, Tennessee, Virginia, Missouri, South Dakota, California, Montana, Minnesota, Arizona
Recall number
Z-2357-2024
Recall status
Open, Classified
CDRH record ID
208517
FDA classified
2024-07-15
Recall event number
94872

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗