Skip to content

Public FDA record

Z-2357-2026: Giraffe Stand-alone Infant Resuscitation System; Powered emergency ventilator (resuscitator) Product iden

Recall record Z-2357-2026 for Datex-Ohmeda Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2357-2026
Company in FDA record
Datex-Ohmeda Inc
FDA classification date
Jun 18, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.

Source facts

Action
Firm began notifying customers on 5/8/2026 via letters titled "URGENT MEDICAL DEVICE CORRECTION." Customers were instructed to identify affected devices by following instructions provided in the letter. Devices with affected blenders should be removed from patient use, and the customer should contact a GE HealthCare Service Representative.
Initiated
2026-05-08
Root cause
Process control
Distribution
Pending
Recall number
Z-2357-2026
Recall status
Open, Classified
CDRH record ID
220638
FDA classified
2026-06-18
Recall event number
99036

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗