Z-2359-2019: MassHunter Quantitative Analysis Software, Version B.07.01, used with Triple Quadrupole Mass Spectrometer

Recall record Z-2359-2019 for Agilent Technologies, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2359-2019
Company in FDA record
Agilent Technologies, Inc.
FDA record creation date
Aug 23, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When utilizing specific software versions of an LC/MS device, and Batch at a Glance in Compound Table view, a defect occurs when a new sample(s) is inserted. This defect creates a mismatch between the sample name and the column header whereby the sample header will be offset by one sample. The recalling firm requests that you discontinue using this workflow to generate reports because the Quant batch table will not display fully analyzed results or save any changes made.

Source facts

Action
On 08/13/18, Urgent Medical Device Recall notices were sent to customers informing them that the defect has been corrected with software version B.08.01. The recalling firm will contact customers to schedule the new software installation. Software version B.08.01 is being released in conjunction with software version B.08.03. This software suite includes improvements to other non-patient safety related defects such as the following fixes in Acquisition B.08.03. -Improvements to DMRM method setup which will prevent exceeding transition limits ("Failed to WriteScan" errors). Improvements to MassHunter Optimizer preventing intermittently missed entries that give "Phase 3Optimization" error.A full list of defects is contained in the Software Release Bulletin supplied with your software. Customers with additional questions were encouraged to call 1-866-488-4080. Customers were asked to complete and return an acknowledgement form.
Initiated
2018-08-15
Root cause
Software design
Terminated
2021-09-28
Distribution
US Nationwide. distribution.
Recall number
Z-2359-2019
Recall status
Terminated
CDRH record ID
173417
FDA record created
2019-08-23
Recall event number
82794

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗