Open original FDA record ↗Recall record
- FDA identifier
- Z-2359-2019
- Company in FDA record
- Agilent Technologies, Inc.
- FDA record creation date
- Aug 23, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
When utilizing specific software versions of an LC/MS device, and Batch at a Glance in Compound Table view, a defect occurs when a new sample(s) is inserted. This defect creates a mismatch between the sample name and the column header whereby the sample header will be offset by one sample. The recalling firm requests that you discontinue using this workflow to generate reports because the Quant batch table will not display fully analyzed results or save any changes made.
Source facts
- Action
- On 08/13/18, Urgent Medical Device Recall notices were sent to customers informing them that the defect has been corrected with software version B.08.01. The recalling firm will contact customers to schedule the new software installation. Software version B.08.01 is being released in conjunction with software version B.08.03. This software suite includes improvements to other non-patient safety related defects such as the following fixes in Acquisition B.08.03. -Improvements to DMRM method setup which will prevent exceeding transition limits ("Failed to WriteScan" errors). Improvements to MassHunter Optimizer preventing intermittently missed entries that give "Phase 3Optimization" error.A full list of defects is contained in the Software Release Bulletin supplied with your software. Customers with additional questions were encouraged to call 1-866-488-4080. Customers were asked to complete and return an acknowledgement form.
- Initiated
- 2018-08-15
- Root cause
- Software design
- Terminated
- 2021-09-28
- Distribution
- US Nationwide. distribution.
- Recall number
- Z-2359-2019
- Recall status
- Terminated
- CDRH record ID
- 173417
- FDA record created
- 2019-08-23
- Recall event number
- 82794