Blender knob shaft can loosen, which can affect the delivered oxygen concentration. The system can deliver an oxygen concentration that does not match the intended concentration set by the user, potentially resulting in hypoxia or hyperoxia in the patient.
Source facts
Action
Firm began notifying customers on 5/8/2026 via letters titled "URGENT MEDICAL DEVICE CORRECTION."
Customers were instructed to identify affected devices by following instructions provided in the letter. Devices with affected blenders should be removed from patient use, and the customer should contact a GE HealthCare Service Representative.
Initiated
2026-05-08
Root cause
Process control
Distribution
Pending
Recall number
Z-2359-2026
Recall status
Open, Classified
CDRH record ID
220640
FDA classified
2026-06-18
Recall event number
99036
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.