Skip to content

Public FDA record

Z-2360-2020: Infant Child Starter Kits, Part Number: 3202784-009, containing Infant/Child Reduced Energy Defibrillati

Recall record Z-2360-2020 for Physio-Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2360-2020
Company in FDA record
Physio-Control, Inc.
FDA record creation date
Jun 11, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Packaging of infant child reduced energy electrodes was not properly sealed.

Source facts

Action
Urgent Medical Device Safety Notice & Correction Action notices dated 04/24/20 were mailed by Stryker to customers. Customers were asked to inspect their inventory to identify any electrode packages that have compromised seals. Remove from inventory and destroy any product suspected to exhibit this condition. Complete the attached acknowledgement form and return the completed form to your distributor to confirm your receipt and understanding of this information. Your distributor will provide you with replacement electrodes. Distributors were asked to communicate recall information to their customers by using an editable Customer communication packet. The End Customer Acknowledgement and Receipt Form should be edited to include remittance information for your organization (recommend email address and/or fax number) in the editable fields at the bottom of the form so that the forms completed by your customers will be returned to you. Upon receipt of End Customer Acknowledgement and Receipt Form from end customers, consolidate information within a copy of the enclosed Acknowledgement and Receipt Form and return to Stryker as directed. Stryker will ship replacement electrodes to distributor upon receipt for you to distribute to your end customers. Customers with additional questions were encouraged to call 1-800-787-9537, option 2, 8:00 A.M. to 6:00 P.M. (Eastern Time), Monday - Friday.
Initiated
2020-04-24
Root cause
Packaging change control
Terminated
2023-07-27
Distribution
Worldwide distribution - US Nationwide distribution including in the states of NY,TX, IL, CO, TN, MI, FL, CT, GA, WA, AZ, NJ, MA, MO, NC, MD, OH, PA, CA, IA, VA, WI, MN, SC, IN, DE, KY, MT, AL, MS, NM, ME, OK, WV, SD, RI, ND, LA, AR, HI, DC, KS, NE, VT, OR, AK, NV, UT, WY, NH, PR, ID, and the countries of CA, ZA, TW, AU, JP, HK, MA, NZ, SG, BR, NL, PA, SA, AR, CL, TH, SV, CR, TR, GU, MY, VI.
Recall number
Z-2360-2020
Recall status
Terminated
CDRH record ID
181540
FDA record created
2020-06-11
Recall event number
85173

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗