Z-2360-2026: Automated Impella Controller (AIC), for use with left heart support blood pump. Labeled as the following

Recall record Z-2360-2026 for Abiomed, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2360-2026
Company in FDA record
Abiomed, Inc.
FDA classification date
Jun 12, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential software error in the Automated Impella Controller ("AIC" ) when used in conjunction with left ventricular Impella devices.

Source facts

Action
Abiomed notified consignees on about 05/14/2026 via letter titled "URGENT MEDICAL DEVICE RECALL (CORRECTION)." Consignees were made aware of the issue and potential risk and were informed that devices were not being removed, and hospital inventory can continue to be used. Specific, outlined scenarios were explained and that if they were to occur, that a console exchange is unnecessary as the phenomenon will recur on the exchange device. Abiomed is actively working on a software update to the AIC to address this issue. Consignees were instructed to complete and return the provided business reply form, forward the notice to anyone in each facility that needs to be informed, post copy of the notice in a visible area for awareness, and if any affected products have been forwarded to another facility, contact that facility and provide them with the notice.
Initiated
2026-05-14
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Austria, Belgium, Brazil, Brunei, Canada, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Luxembourg, Malaysia, Mexico, Netherlands, New Zealand, Norway, Panama, Poland, Saudi Arabia, Serbia, Singapore, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom.
Recall number
Z-2360-2026
Recall status
Open, Classified
CDRH record ID
219986
FDA classified
2026-06-12
Recall event number
98853

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗