Z-2361-2020: Piccolo Xpress chemistry analyzer, Model No. 1100-1001, Software version 2.1.55 (NGA)/3.1.35(Jen II)

Recall record Z-2361-2020 for Abaxis Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2361-2020
Company in FDA record
Abaxis Inc
FDA record creation date
Jun 12, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Incorrect reference ranges of analytes.

Source facts

Action
On January 9, 2020, the firm sent an "Urgent Medical Device Recall' to their customers, regarding the chemistry analyzer containing software versions software versions 2.1.55/3.1.3. 5. This software failure may cause the product to display the incorrect reference ranges for certain analytes when gender selection is not set on the device. Erroneous reference values may result in an incorrect diagnosis or treatment plan.The firm informed customers to immediately stop using the devices and return it to the firm. On January 21, 2020, the firm instructed their customers to check the enclosed list of affected analytes and their correct reference range values and informed them that the erroneous reference values may result in an incorrect diagnosis or treatment plan. The affected products are shipped between 11/29/19 and 12/3/19, as either a unit on loan or as a repair/service of unit. The firm instructed customers to immediately examine their inventory and quarantine the product subject to recall. The firm ceased further distribution or use of the product. The firm instructed to the customers that their Technical Support arranged for the product return to the firm and would provide the repaired or replacement at no cost. The firm provided the Technical Support number as 1-800-822-2947 for customers to contact them.
Initiated
2019-12-11
Root cause
Software design
Terminated
2022-07-22
Distribution
All U.S. Consignees, TX, PA, FL, OK, KY, CA, VA No OUS Consignees.
Recall number
Z-2361-2020
Recall status
Terminated
CDRH record ID
181170
FDA record created
2020-06-12
Recall event number
85533

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗