Skip to content

Public FDA record

Z-2362-2008: Boston Scientific Platinum Plus Guidewire. Material No. [outer carton label] H74917511, Catalog # 1751,

Recall record Z-2362-2008 for Boston Scientific Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2362-2008
Company in FDA record
Boston Scientific Corporation
FDA classification date
Sep 21, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Boston Scientific is initiating a voluntary recall of the Platinum Plus Coronary Guidewire. Through our internal inspection process, we identified the potential that the PTFE (polytetrafluoroethylene) coating on the Platinum Plus Coronary Guidewires could be damaged in certain locations. This damage of the coating creates the potential for small particles of the PTFE coating to detach from the wi

Source facts

Action
An urgent Voluntary medical Device Recall or Correction - Immediate Action Required letter was sent to all affected customers. The letter describes the issue and clinical effects. A list of affected devices is attached to the letter. The customer is to cease use or distribution of affected product, and return it immediately to Boston Scientific. Return directions are provided in an attached document. A Reply Verification Tracking Form is to be completed and faxed back to Customer Service. If any product is being returned the completed form is to accompany the returned product. Contact Boston Scientific Corporation at 1-763-494-1700 for assistance.
Initiated
2008-06-19
Root cause
Process control
Terminated
2010-10-13
Distribution
AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, ME, MI, MN,MO, MS, MT, NC, ND, NE, NH, NJ, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, SI, WV, and WY. OUS locations to include: Balkans, Finland, Germany, Great Britain, Hong Kong, Ireland, Israel, Italy, Mexico, Netherlands, Norway, Romania, Saudia Arabia, Spain, Sweden, Switzerland, and United Arab Emirates
Recall number
Z-2362-2008
Recall status
Terminated
CDRH record ID
72494
FDA classified
2008-09-21
Recall event number
49001

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗