Z-2362-2026: Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1

Recall record Z-2362-2026 for The Binding Site Group, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2362-2026
Company in FDA record
The Binding Site Group, Ltd.
FDA classification date
Jun 9, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.

Source facts

Action
Firm notified customers via "Urgent Medical Device Correction" letter dated April 14, 2026. Customers were advised to confirm QC results are present in the QC results screen for the corresponding plate barcode after the completion of each plate measurement. If no QC result is displayed, do not release results from that plate and repeat the run ensuring QC material is properly loaded. The customer should also contact US Technical Support. When performing multi-plate runs while loading a previously used QC vials, ensure newly opened QCs are loaded onto Reagent Runner 2 as well, or discard and load newly opened QC vials into Reagent Runner 1. Retrospective review of previously released results is not considered necessary by the recalling firm. Please maintain awareness of the issue until Thermo Fisher Scientific identifies and implements a software enhancement that addresses the issue.
Initiated
2026-04-16
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution.
Recall number
Z-2362-2026
Recall status
Open, Classified
CDRH record ID
219899
FDA classified
2026-06-09
Recall event number
98830

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗