Skip to content

Public FDA record

Z-2363-2019: Alinity ci -series System ControlModule, Part Number 03R70 01 Product Usage: The Alinity ci series Syste

Recall record Z-2363-2019 for Abbott Gmbh & Co. KG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2363-2019
Company in FDA record
Abbott Gmbh & Co. KG
FDA record creation date
Aug 24, 2019
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Abbott has identified an issue with all on market versions (v2.6.0 and v2.6.1) of Alinity ci series Software where incorrect results may occur after a system Stop due to the Alinity i re use of reaction vessels (RVs). This issue only occurs if the system is transitioned from Processing to Stopped to Idle.

Source facts

Action
Abbott sent an Urgent Product Correction Notice by letter dated 06/10/2019. The letter identified the affected product, problem and actions to be taken. Abbott has identified an issue with all onmarket versions of Alinity ciseries Software where reuse of reaction vessels (RVs) may occur after a system Stop. This issue only occurs if the system is transitioned from Processing to Stopped to Idle and has the potential to cause incorrect results.The letter provided a mitigation strategy for the consignee to use until the updated software is available. For questions U.S. Customers contact Customer Service at 18774ABBOTT (available 24 hours a day, 7 days a week). Customers outside the U.S., please contact your local area Customer Service.
Initiated
2019-06-10
Root cause
Software in the Use Environment
Terminated
2023-06-14
Distribution
Worldwide Distribution - US Nationwide
Recall number
Z-2363-2019
Recall status
Terminated
CDRH record ID
174125
FDA record created
2019-08-24
Recall event number
83235

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗