Skip to content

Public FDA record

Z-2364-2010: 22G x 70mm/20 bevel Single Shot Needle, REF 552-022-70, Rx, Sterile, Distributed by: Stryker Instrumen

Recall record Z-2364-2010 for Stryker Instruments Div. of Stryker Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2364-2010
Company in FDA record
Stryker Instruments Div. of Stryker Corporation
FDA classification date
Sep 3, 2010
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Defects include, but not limited to: Not sterile, not properly labeled, drugs and swabs may be ineffective, and catheter tips may break in patients.

Source facts

Action
Stryker Instruments issued an "Urgent Medical Device Recall Notification" letter to consignee risk managers, directors and material managers on February 2, 2010, identifying the affected devices and product issues. The consignees were instructed to: 1) Forward the recall notice and reply form to all affected locations. 2) Review all inventory and quarantine affected product. 3) Destroy all affected product using attached instructions. 4) Complete and return the Business Reply Form to the firm by fax. Stryker Instruments can be contacted at 1-9800-800-4236 ext 4354
Initiated
2010-02-01
Root cause
Process control
Terminated
2011-03-28
Distribution
Nationwide Distribution: United States, including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, TN, TX, and UT.
Recall number
Z-2364-2010
Recall status
Terminated
CDRH record ID
88813
FDA classified
2010-09-03
Recall event number
54625

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗