Open original FDA record ↗Recall record
- FDA identifier
- Z-2364-2019
- Company in FDA record
- CooperSurgical, Inc.
- FDA record creation date
- Aug 24, 2019
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm updated the IFU in April 2019 to clarify that use of the TRANSWARMER Mattress with other heat producing devices, such as an incubator, is prohibited and could lead to serious health consequences, such as skin burns. some products do not contain the updated IFU.
Source facts
- Action
- On July 8, 2019, the firm began notifying customers of the recall via Urgent Medical Device Correction letters. The letters informed customers that CooperSurgical has updated the IFU to clarify that use of the Transwarmer Mattress with other heat producing devices, such as an incubator, is prohibited. Such use could lead to serious health consequences, such as skin burns. Customers were provided with an updated version of the IFU. Customers were asked to complete and return the acknowledgement and receipt form to the firm. If you have any questions about this recall, please contact CooperSurgical at 203-601-5200 ext. 3300.
- Initiated
- 2019-07-08
- Root cause
- Error in labeling
- Terminated
- 2020-05-28
- Distribution
- Worldwide Distribution - US Nationwide in all 50 states. International distribution to Australia, Brazil, Canada, Ecuador, Germany, Gibraltar, India, Ireland, Kuwait, Macedonia, New Zealand, Oman, Poland, Puerto Rico, Qatar, Romania, Saudi Arabia, South Korea, Switzerland, United Arab Emirates, & United Kingdom.
- Recall number
- Z-2364-2019
- Recall status
- Terminated
- CDRH record ID
- 173725
- FDA record created
- 2019-08-24
- Recall event number
- 83110