Z-2364-2023: UroLift Visual Obturator, 20Ch-The visual obturator is used with the UroLift system. Instrument for inser

Recall record Z-2364-2023 for Scholly Fiberoptic Gmbh. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2364-2023
Company in FDA record
Scholly Fiberoptic Gmbh
FDA classification date
Aug 9, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is an improperly performed weld between the cone of the sheath lock and tube of the obturator; mechanical load capacity is impaired, and fracture of the welded joint may occur. Contamination can occur in this capillary gap, which is not covered by the reprocessing validations performed.

Source facts

Action
SCHOLLY FIBEROPTIC GMBH initiated Recall Plan Letter to Users, Importers and Distributors . The US Distributor- Teleflex (parent company of NeoTract) notified on June 23, 2023 to forward the notification letter to their customers. Letter states reason for recall, health risk and action to take: Review this Field Safety Notice in its entirety and ensure that all users of the above-mentioned product in your organization and other concerned persons are informed about this Field Safety Notice. 2. If you are a distributor, please forward this correction notification to your customers. 3. Identify, quarantine and return affected devices. 4. Do not use affected devices anymore. 5. Exchange devices from the above-mentioned batches, which are currently in use. 6. Confirm receipt of this information by filling in the attached reply form and returning it to the indicated e-mail address. E-Mail: vigilance@schoelly.de
Initiated
2023-06-23
Root cause
Process control
Distribution
Nationwide Foreign: Japan, South Korea
Recall number
Z-2364-2023
Recall status
Open, Classified
CDRH record ID
202206
FDA classified
2023-08-09
Recall event number
92700

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗