During regular product monitoring, firm became aware of a potential issue with the Mobilett Mira system, (serial numbers 1034, 1038, 1041, 1042, 1043, 1050, 1053, and 1054. During normal clinical operation the center screw on the driving wheel may become loose resulting in system immobilization. There is no risk of the system tipping over or collapsing on the floor, however the system may tilt t
Source facts
Action
Siemens has issued an Update Instruction XP018/12S in April 2012 to all affected consignees notifying them of the problem. The update instruction provides a correction of the wheel fixation. For questions call Siemens at 610-219-4834 or 610-448-4634.
Initiated
2012-04-30
Root cause
Device Design
Terminated
2012-09-25
Distribution
US Nationwide Distribution including the states of: GA, IA, KS, NC and TX.
Recall number
Z-2367-2012
Recall status
Terminated
CDRH record ID
112211
FDA classified
2012-09-11
Recall event number
62989