Skip to content

Public FDA record

Z-2369-2008: Philips Telemonitoring System Software (M3811B) used with Philips Telemonitoring System (M3810B) Revisio

Recall record Z-2369-2008 for Philips Medical Systems. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2369-2008
Company in FDA record
Philips Medical Systems
FDA classification date
Sep 18, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Software: If Weight Limits edited, values will return to default values that were deleted

Source facts

Action
Affected customers will receive the Urgent Medical Device Correction notification on 7/22/08 via certified mail. Delivery receipt will be tracked. The Urgent Medical Device Correction notification instructs customers to verify that daily and weekly weight limits are set or not set as intended while they await the correction, which will come in the form of a new version of software and will simultaneously update all systems. Philips has corrected the software defect and is in the process of validating the correction and releasing the corrected software as part of a scheduled release. If further information or support is needed, customers should contact the Telehealth Customer Service Line at 1-866-246-7316.
Initiated
2008-07-22
Root cause
Software design
Terminated
2012-05-16
Distribution
Nationwide Distribution.
Recall number
Z-2369-2008
Recall status
Terminated
CDRH record ID
72508
FDA classified
2008-09-18
Recall event number
49005

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗