Skip to content

Public FDA record

Z-2370-2026: Description/REF: PI AGBA PICC KIT: 3L 6 FR X 40 CM W/VPS/ASK-44063-NS AGBA PICC/DELTA: 2L 5.5 FR X 50 CM/

Recall record Z-2370-2026 for ARROW INTERNATIONAL, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2370-2026
Company in FDA record
ARROW INTERNATIONAL, LLC
FDA classification date
Jun 13, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Lidocaine, Bupivacaine, and 0.9% sodium chloride, the subject of a supplier recall, were included into kits and sets. The supplier was found to have quality issues and limited reports of drug ineffectiveness, and if sterility of a water-based injectable product is compromised this could lead to serious or life-threatening injuries.

Source facts

Action
On 4/21/2026, correction notices (Firm Event Number: EIF-000608, lidocaine and bupivacaine) were mailed to customers who were asked to do the following: Arrow International LLC is voluntarily recalling product due to the Huons Co. Ltd. recall of Lidocaine and Bupivacaine, which is included in impacted Arrow Kits & Sets. Medical facilities 1. Provide this correction notice to all persons who need to be aware within your organization or to any organization where the potentially affected product has been transferred. 2. Check your inventory for product within the scope of this Medical Device Notification and place a copy of this notice with all affected product. 3. Complete and return the response form virtually via: https://fca.myteleflex.com/en/recall/000608 Distributors: 1. Provide Medical Device Notification to all customers who have received impacted product. Each of your customers is then required to complete the Acknowledgement Form and return it to you as their distributor. 2. Check your inventory for affected product. Place a copy of this notification with all affected product and ensure that the notification is included with future distributions of affected product. 3. Complete and return the response form via link: https://fca.myteleflex.com/en/recall/000608 4. If you have further distributed product outside of your country, please notify Teleflex Customer Service by return email. If you have questions contact firm's Customer Service: Link: https://fca.myteleflex.com/en/recall/000608, Customer Service: Telephone: 1-866-396-2111, FAX: 1-855-419-8507 Email: Recalls@teleflex.com On 5/29/2026, additional amended notices (Firm Event Number: EIF-000608-01) were mailed to include additional affected devices. On 5/1/2026, additional notices (Firm Event Number: EIF-000609, sodium chloride) were mailed and emailed to customers who have 0.9% sodium chloride (10mL saline solution ampule) in kits and sets.
Initiated
2026-04-21
Root cause
Nonconforming Material/Component
Distribution
US Nationwide distribution including in the states of VA, NY, UT, SC, LA, MN, PA, AZ, CA, MI, OH, FL, CT, KS, IL, TX, WI, GA, NC, MD, OR, AK, NH, IA, ND, OK, NM, AL, MA, KY, WA, AR, CO, IN, ME, TN, WY, NJ, SD, MS, MO, MT, DE, ID, NE, WV, RI, VT, NV, HI, DC, PR.
Recall number
Z-2370-2026
Recall status
Open, Classified
CDRH record ID
220062
FDA classified
2026-06-13
Recall event number
98793

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗