Skip to content

Public FDA record

Z-2371-2020: PENTARAY NAV HIGH-DENSITY MAPPING CATHETER ELECTRODE EA 22 Rx only STERILE EO - Product Usage: The B

Recall record Z-2371-2020 for BIOSENSE WEBSTER. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2371-2020
Company in FDA record
BIOSENSE WEBSTER
FDA record creation date
Jun 16, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Mapping Catheter can become entrapped in the cardiac valvular apparatus while the user is attempting to create an electroanatomic map.

Source facts

Action
On 4/22/2020 the firm issued an "URGENT Medical Device Correction" letter via UPS/certified mail, informing them that as a result of 3 complaints in 2019 in which a catheter became entrapped in the cardiac valvular apparatus while the user was attempting to create an electroanatomic map, leading to the need for surgical intervention for removal, the firm is performing a correction to the products. The Notification Letter informs customers: Following are learnings that users of this product should be aware of to avoid or address entrapment: " Repeated rotation of the catheter near the valvular apparatus may result in the entanglement of the product spines and further rotation could result in entrapment, leading to the need for surgical intervention. " A common use scenario for repeated rotation leading to entanglement may occur during mapping of the right ventricle and accessing of the right ventricular outflow tract using clockwise rotation. Caution should be exercised with manipulation to avoid spine entanglement with valvular structures. " Entanglement of the catheter can be identified by resistance felt by the user and a lack of movement of the spines during manipulation, which can be seen via the CARTO 3 System or by fluoroscopy. The device may appear in a broom-like configuration with entangled spines pointing in a single direction. With opposite manipulation (e.g. counterclockwise rotation for initial clockwise rotation), it may be possible to free the entangled device for safe removal. " Do not use excessive force to remove the device when entangled or entrapped as it may lead to damage to the cardiac structures or to the device. The firm is instructing customers: 1. Please maintain awareness of this letter and pass it on to anyone in your facility that needs to be informed. 2. Please complete, sign, and return the Business Reply Form. We have communicated this information to the applicable regulatory authorities. Method of Notification/M
Initiated
2020-04-22
Root cause
Error in labeling
Terminated
2022-12-15
Distribution
Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, FL, GA, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NH, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI IL, VA,, NE, OK, NJ, CT, KS, HI, CO, ND, RI, ID, MS, VT, DC and WV. The countries of (EMEA, APAC, LATAM, US, Canada) including: NZ,BE,DE,PL,SI,TR,HU,CO,KR,CL,PE,AU,BR,CA,CN,HK,ID,JP,MY,MX,PH,RU,SG,TW,TH,IL,AT,FR,GR,DE,IE,IT,PL,NL,PT,ES,SE,DK,NO,FI,CH,GB,RO,SI,HR,CZ,TR,ZA,AE,BR,ID,BE,PA,CR,CL,NZ,HK,RU,GR,NL,EG,QA,LU,NL,CH,AT,IE,CY. See Attachments 2 & 3 for a distribution list including all domestic and foreign consignees of the device, addresses and phone numbers and the dates and number of devices distributed to each such consignee.
Recall number
Z-2371-2020
Recall status
Terminated
CDRH record ID
181237
FDA record created
2020-06-16
Recall event number
85559

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗