Open original FDA record ↗Recall record
- FDA identifier
- Z-2373-2015
- Company in FDA record
- Fresenius Medical Care Holdings, Inc.
- FDA classification date
- Aug 12, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Expansion of recall to lots manufactured in 2015 due to the potential for bacterial contamination.
Source facts
- Action
- Fresenius Medical NA issued an Urgent Expanded Medical Device Recall letter on 7/16/15 to clinics and patients. The letter identified the affected product, explained the reason for the expansion, and instructed users to immediately examine their stock to determine whether there was any Naturalyte Liquid Bicarbonate Concentrate beginning with 15AMLB, 15BMLB, 15CMLB. Users are to discontinue use immediately if any product of these lots is found, . All units are to be placed in a secure, segregated area. If affected product was on the machine prior to patient treatment, a [Heat Disinfect] program should be performed. The attached Reply Form should be completed and returned. Contact your FMCNA Customer Service Team at 1-800-323-5188 for instructions on how to return the recalled product.
- Initiated
- 2015-07-16
- Root cause
- Process control
- Terminated
- 2026-11-03
- Distribution
- Nationwide Distribution
- Recall number
- Z-2373-2015
- Recall status
- Terminated
- CDRH record ID
- 138829
- FDA classified
- 2015-08-12
- Recall event number
- 71724