Z-2373-2024: Horizon X-Ray Bone Densitometer (DXA) Models: Horizon-A, Horizon-W, Horizon-WI, Horizon-C, Horizon-CI, H

Recall record Z-2373-2024 for Hologic Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2373-2024
Company in FDA record
Hologic Inc
FDA classification date
Jul 16, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Non-conformance pertains to electromagnetic compatibility requirements according to the international technical standard IEC 60601-1-2 for the safety and essential performance of medical electrical equipment, where the result from the Horizon DXA System exceeded the limit for acceptability.

Source facts

Action
***August 16, 2024 UPDATE*** After further review of the install base for systems in scope of the Horizon recall, updates made to the consignee list. From the original list, 270 systems have been identified to be out of scope of this recall. Additionally, 44 systems were identified to be in scope of the recall that were not on the original consignee list. Customer notifications for the additional 44 in scope systems were distributed on 8/13/2024. RES Updated Serial numbers. Hologic issued URGENT: MEDICAL DEVICE CORRECTION FA-00247 Letter on 6/26/24. Letter states reason for recall, health risk and action to take: Recommendations: " Do NOT scan patients that have active implanted medical devices, including but not limited to neurostimulators, pacemakers, cardiac defibrillators, continuous glucose monitors, or other bio-wearable sensors. " Any operator who has an active implanted medical device should also refrain from operating the system at this time. " Do NOT scan patients that are currently being treated with an electronic medical device. " Extend this communication to all pertinent staff in interaction and/or use with the Horizon DXA System. " Until the correction is completed, this letter and the specific warning below supersedes information provided on the Horizon DXA labeling and IFU pertaining to electromagnetic compatibility and electromagnetic interference. Rectification activity by Hologic: We are urgently investigating the permanent rectification actions required, and we will be in contact promptly once this is defined. A service appointment will be scheduled for the remediation activity. As additional clarification regarding the continued safe use of your Horizon DXA system, please note the following: This notice pertains to electromagnetic emissions, not ionizing radiation. Electromagnetic emissions are emitted by all electronic devices, such as cell phones, lights, computers, TVs, medical devices, among others. Further, please note that elec
Initiated
2024-06-13
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of India, Australia, Spain, Estonia, Germany, Italy, Canada, Thailand, Dominican Republic, Portugal, Islas Canarias, Andorra, Switzerland, Taiwan, Austria, Korea, Republic of Philippines, Kuwait, Slovakia, China, Belgium, Ireland, United Kingdom, Malaysia, Palestinian Territory, Occupied, Peru, Singapore, Qatar, Algeria, Pakistan, New Zealand, Croatia, Brazil, Bangladesh, Saudi Arabia, Venezuela, Israel, Lebanon, Nepal, Romania, Greece, Hong Kong, Morocco, Indonesia, Cyprus, Jordan, France, Mexico, Colombia, Iran, Islamic, Republic of Chile, Turkey, Uruguay, Viet Nam, Serbia, Poland, Japan, Malta, Netherlands, South Africa.
Recall number
Z-2373-2024
Recall status
Open, Classified
CDRH record ID
208749
FDA classified
2024-07-16
Recall event number
94905

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗