- FDA identifier
- Z-2374-2016
- Company in FDA record
- Teleflex Medical
- FDA classification date
- Aug 9, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Cracks may occur at the 15mm and 22mm ISO female connector.
Source facts
- Action
- Inmed, a division of Teleflex sent an Urgent Field Safety Notice letter dated December 3, 2015 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to check inventory, cease use and distribution of stock of the affected product batch and quarantine immediately.. The letter included an acknowledgment form which was to be completed and returned your MAQUET representative. For further information or support concerning this issue, contact your local MAQUET representative at fieldactions@maquet.com. For questions contact your local sales representative or Customer Service.
- Initiated
- 2015-12-03
- Root cause
- Release of Material/Component prior to receiving test results
- Terminated
- 2017-04-14
- Distribution
- US Nationwide Distribution in the states of FL, KS and WV.
- Recall number
- Z-2374-2016
- Recall status
- Terminated
- CDRH record ID
- 148176
- FDA classified
- 2016-08-09
- Recall event number
- 74681