Z-2374-2024: BD Pyxis: MedStation ES (Main), REF: 323; MedStation ES Tower, REF: 352; Anesthesia Station ES, REF: 327;

Recall record Z-2374-2024 for CareFusion 303, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2374-2024
Company in FDA record
CareFusion 303, Inc.
FDA classification date
Jul 17, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

If automated dispensing cabinets have specific software versions, and Component Manager was configured to "installed mode", and software patch KB 5033688 was installed in the production environment, then this will will lead to an error that prevents the user from accessing the dispensing software application, which may delay accessing medication.

Source facts

Action
On 6/6/24, correction notices were distributed to customers who were informed that their devices had been remediated. That they had been contacted by phone in May 2024 to schedule device remediation. The firm re-configured the component manager from "installed mode" to "disabled mode" and checked if the Microsoft patch was present on each device. If the patch was identified on a device, it was removed. All devices were then configured to "managed mode" and rebooted to enable the system appropriately. Customers are asked to complete and return the response form to Email BDRC15@bd.com For safety information about the system not covered in this document, call the firm's Technical Support Center at 1-800-727-6102, Phone hours: 24 hours, 7 days a week. Additional contacts include: 1) Regional Complaint Center, 1-844-823-5433 Say "Product Complaints" when prompted Mon-Fri 8:00am and 5:00pm CT, Email: productcomplaints@bd.com 2) Post Market Quality Field Action Operations, BDRC15@bd.com
Initiated
2024-05-02
Root cause
Software change control
Distribution
Worldwide - US Nationwide distribution including in the states of OR, KS, NE, LA, ND, OK, AL, AK, WI, WV, WA, NV, PR, TX, CO, OH, CA, MO, IA, AZ, IN, AR, IL, MN, HI, NM, FL, SD, UT, NY, VA, GA, MA, NC and the countries of TW, CA.
Recall number
Z-2374-2024
Recall status
Open, Classified
CDRH record ID
208032
FDA classified
2024-07-17
Recall event number
94699

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗