Z-2376-2021: NORMOFLO Irrigation Fluid Warmer Models H-1100 and H-1129

Recall record Z-2376-2021 for Smiths Medical ASD Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2376-2021
Company in FDA record
Smiths Medical ASD Inc.
FDA record creation date
Sep 3, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

There is potential for harmful levels of aluminum leaching when using lower flow rates with certain solutions and blood products and longer duration of use.

Source facts

Action
Smiths Medical sent an Urgent Medical Device Recall Notice and attachments dated August 13, 2021 to all affected customers. The letter instructed all customers who purchased affected devices listed in the table on page 1 of the Urgent Medical Device Recall Notice must identify any of the affected products within their possession and refer to the detailed instructions on pages 2 and 3 of the notice. Distributors will be instructed to immediately forward a copy of the Urgent Medical Device Recall Notice and attachments to any customers to whom they have distributed affected product. Distributors will also be instructed to request that customers to who they have distributed affected product complete the Response Form and return it to the distributor. Distributors will indicate their identity as the distributor and document the consignees name and address. Questions regarding this recall notification may be forwarded to fieldactions@smiths-medical.com.
Initiated
2021-08-06
Root cause
Device Design
Distribution
US Nationwide Distribution and PR OUS distribution in - Austria, Australia, Belgium, Switzerland, Canada, Germany, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Hungary, Ireland, Italy, Kuwait, Luxembourg, Netherlands, New Zealand, Panama, Poland, Portugal, Sweden, Turkey, Venezuela, South Africa
Recall number
Z-2376-2021
Recall status
Open, Classified
CDRH record ID
188709
FDA record created
2021-09-03
Recall event number
88455

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗