Z-2378-2012: The product affected by this notice is sold under the trade name TomoTherapy SharePlan 1.1. To determine

Recall record Z-2378-2012 for RAYSEARCH LABORATORIES AB. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2378-2012
Company in FDA record
RAYSEARCH LABORATORIES AB
FDA classification date
Sep 14, 2012
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 17, 2026
Recall classification
Class II

Reason stated in the source

This field notice concerns two behaviors of the SharePlan (Multiple Fraction Groups and Depth Offset) that may be unexpected and are not clearly described in the user manual. None of these behaviors have caused patient mistreatment or other incidents. However, the described user actions must be adhered to for best agreement between treatment plan evaluation and the clinical outcome of the delivere

Source facts

Action
RaySearch Laboratories sent an Urgent Field Safety Notice Medical Device Correction dated May 30, 2011, to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consigness were instructed to completely avoid affecting the erroneous behaviour, verify in TomoTherapy Hi-Art that the plans sent to TomoTherapy SharePlan 1.1 only have one single fraction group with a single fraction size,used for all fractions, and do not attempt to import treatment plans having more than one fraction size into the product. To completed avoid affecting the erroneous behaviour the user can either offset the measured curves before importing them into beam commissioning or add the depth offset tothe calibration depth. For further information consigness should contact RaySearch Laboratories AB Quality and Regulatory Affairs at +46 722 366 110 or the Tomo Therapy support at support@tomotherapy.com. North America 1 866 368 4807 Belgium 0800 38783 France 0805 631 565 Germany 0800 000 +541 Italy 800-986 399 Netherlands 0800 0201364 Spain 800 3000049 Switzerland 0800 001927 UK 0868 238 6035 China/Northern 10 800 712 1701 China/Southern 10800 120 1701 Hong Kong 800 967912 Japan 0044 22 132374 Singapore 800 1204683 South Korea 0079 81 4800 7204 All other locations + 1 608 824 2900 or +32 2 40 44 44
Initiated
2011-05-30
Root cause
Labeling False and Misleading
Terminated
2012-11-16
Distribution
Worldwide Distribution - USA including CA, KY, MI, NY, OK, and OR. Internationally to Canada, Belgium, United Kingdom, Australia, France, Germany, and Poland.
Recall number
Z-2378-2012
Recall status
Terminated
CDRH record ID
112573
FDA classified
2012-09-14
Recall event number
63110

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗