Products labeled as sterile were distributed, but may not have been sterilized. Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter Data was manipulated to appear to be conforming
Source facts
Action
The firm notified its consignees of the recall by letter on 07/22/2021. The letter explained the issue and requested the return of the affected product. Distributors were directed to notified their consignees.
Initiated
2021-07-22
Root cause
Vendor change control
Terminated
2024-01-17
Distribution
US Nationwide distribution.
Recall number
Z-2379-2021
Recall status
Terminated
CDRH record ID
188475
FDA record created
2021-08-27
Recall event number
88351