Skip to content

Public FDA record

Z-2380-2018: da Vinci Xi EndoWrist Stapler 45 Reload Blue, Model 48645B-03, 12 reloads/box.

Recall record Z-2380-2018 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2380-2018
Company in FDA record
Intuitive Surgical, Inc.
FDA record creation date
Jul 6, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Intuitive Surgical, Inc. (ISI) has recently become aware of an issue with the da Vinci Xi Stapler 45 Blue and Green Reloads that may be missing a component of the knife mechanism. Without this component, the reload knife will not move and cut tissue when the stapler is fired.

Source facts

Action
Intuitive Surgical, Inc. (ISI) conducted the recall and notified customers via letter "Field Safety Notice Urgent Medical Device Notification" sent the week of April 23, 2018. The letter notified customers of the recall, reason for field action, risk to health, affected products, Actions to be taken by the Customer/User, Actions to be taken by Intuitive Surgical, and Further Information and Support: If you need further information or support concerning this Medical Device Notification , please contact your Clinical Sales Representative or contact Intuitive Surgical Customer Service at the numbers listed below: "North and South America: (800) 876-1310, Option 3 (4 AM to 5 PM PST) or mail: customerservice@intusurg.com. "Europe, Middle East, Asia and Africa: +800 0821 2020 or +41 21 821 2020 (8AM to 6 PM CET) or ics@intusurg.com "South Korea: 02-3271-3200 (9 AM to 6 PM KSTJ) "Japan: 0120-56-5635 or 03-5575-1362 (9 AM to 6 PM JST).
Initiated
2018-04-23
Root cause
Nonconforming Material/Component
Terminated
2020-07-15
Distribution
U.S., foreign and government accounts. Foreign: Australia, Austria, Belgium, Brazil, Canada, China (Hong Kong), Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Qatar, Saudi Arabia, Singapore, South Korea, Spain, Sweden, Switzerland, Taiwan, Turkey, United Kingdom.
Recall number
Z-2380-2018
Recall status
Terminated
CDRH record ID
165193
FDA record created
2018-07-06
Recall event number
80318

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗