Z-2380-2021: Gyrus Acmi PKS Omni 5mm/33cm-intended for the electrosurgical coagulation, mechanical grasping, and disse

Recall record Z-2380-2021 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2380-2021
Company in FDA record
Olympus Corporation of the Americas
FDA record creation date
Aug 27, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Expired adhesive used as an electrical insulation barrier and as a secondary retention bond failed and could result in foreign body / biologic reaction within the patient and / or electrical continuity to patient causing tissue burn to critical structures.

Source facts

Action
Olympus issued Urgent Medical Device Recall Letter on June 9th 2021 via Federal Express two day mail. Letter states reason for recall, health risk and action to take: 1. Immediately cease any further use of any affected product you have, remove it from your inventory and quarantine it until it is shipped back to us. 2. Call your Olympus customer service representative at 1-888-524-7266 option 1. Olympus will issue a Return Material Authorization to return any affected product. at no charge to you. Olympus will issue a credit or replacement to your facility for your affected product. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number Recall-0395 and provide your contact information as indicated in the portal. In addition, if you have further distributed this product, please identify your customers, notify them at once of this product recall, and appropriately document your notification process. Your notification to your customers may be enhanced by including a copy of this recall notification letter. Contact me directly at 484-896-5688 or at laura.storms@olympus.com for any additional information concerning this matter.
Initiated
2021-06-09
Root cause
Process control
Terminated
2023-02-17
Distribution
US Nationwide distribution in the states of CA, GA, MN, TN.
Recall number
Z-2380-2021
Recall status
Terminated
CDRH record ID
188700
FDA record created
2021-08-27
Recall event number
88448

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗