The system is designed to emit a beep upon termination of an exposure. However, if the system has been powered on for over 12 hours, the system will no longer emit this signal due to a defect in the Microsoft Windows 7 operating system.
Source facts
Action
Philips Healthcare sent an Urgent Field Safety Notice letter dated June 25, 2015 to customers. The letter identified the affected product, problem and actions to be taken. Philips plans to install a software upgrade in affected systems, which will eliminate this issue. Philips Healthcare Field Service Engineers will schedule an appointment with customers to install the software update. For questions contact your local philips representative.
Initiated
2015-03-23
Root cause
Radiation Control for Health and Safety Act
Terminated
2017-01-17
Distribution
US Nationwide Distribution to the states of : Arizona, Minnesota, State of Washington, Wyoming and Ohio. Internationally to Canada.
Recall number
Z-2383-2015
Recall status
Terminated
CDRH record ID
139186
FDA record created
2015-11-11
Recall event number
71838