Z-2383-2021: CoolSculpting Elite System The CoolSculpting System SW release 2.0, SW release 3.0, SW release 3.0.1, SW

Recall record Z-2383-2021 for Zeltiq Aesthetics, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2383-2021
Company in FDA record
Zeltiq Aesthetics, Inc
FDA classification date
Aug 27, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

An incorrect error messaging system that could potentially lead to: 1) Reporting a thermal event error causing a user to re-treating the affected anatomic area within 24 hours, 2) Not reporting a thermal event or any other error codes causing a user to continue treating without being aware that a thermal event has occurred.

Source facts

Action
On 07/13/2021 and 7/23/2021, the firm sent "URGENT" product correction notifications to customers informing that software updates were needed to correct error messaging bugs. As a result of the bugs thermal events 1) may not lead to "thermal event" error message alert and treatment would not be stopped; or 2) the error text displayed may be unrelated and the provider would not know to avoid retreatment within 24 hours. These could result in cold-induced injury and 2nd or 3rd degree freeze burns. No injuries have been reported. The firm also informed customers that a corrective software update was deployed remotely to all units and 1) to initiate the update by performing a soft power cycle from the touchscreen; 2) to be aware that the 5 minute process could not be delayed or deferred; 3) not to power off the device during update; 4) to restart the device after update; 5) to verify the successful installation by pressing TOOLS ABOUT, scroll to column Software Release, which should show 3.1.0; and 6) that the software update must be completed before continuing to use the system. If customers have not received the notification of the update on the System or the update did not complete, they should contact product support at 1-888-935-8471 or email at CoolSculpting.support@allergan.com.*** ***Updated/Revised: On 08/26/2021, the firm sent an update/revised "URGENT MEDICAL DEVICE CORRECTION" Notification Letter via email informing its customers to cease use of the CoolSculpting Elite Devices until further notice. The updated communication instructed customers to refrain from using affected devices until the Recalling Firm notifies them because the software change needs to be submitted to FDA for review. For further assistance contact Product Support at 1-888-935-8471 (6am to 5pm PST) or CoolSculpting.support@allergan.com.
Initiated
2021-07-05
Root cause
Software design
Terminated
2023-11-09
Distribution
U.S.: AL, AR, AZ, CA, CO, CT, FL, GA,HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Italy, Norway, Sweden, France, Netherlands, Germany, Spain, United Kingdom, Portugal, South Korea, Singapore, and Hong Kong.
Recall number
Z-2383-2021
Recall status
Terminated
CDRH record ID
188569
FDA classified
2021-08-27
Recall event number
88397

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗