- FDA identifier
- Z-2384-2010
- Company in FDA record
- Whatman, Inc.
- FDA classification date
- Sep 7, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Neonate newborn screening card demographic and filter portion may be detached
Source facts
- Action
- The firm, Whatman, issued an "URGENT MEDICAL DEVICE CORRECTION" letter dated October 29, 2009, to the customer. The letter described the product, problem and action to be taken by the customer. The letter instructed the customer to: inspect the card and ensure the 903 Paper Portion is firmly and completely fixed to the Demographic portio, prior to recording patient demographic details on the card and obtaining a patient sample; discard the card, obtain a new card and inspect as above, if the 903 Paper Portion of the card partially or completely detaches from the Demographic portion of the card; and to please staple the card stub at two points as specified in the illustration, to prevent any opportunity for 903 paper detachment from the demographic portions. If Whatman customers/users have any questions in relation to this customer notification, please contact, either Product Manager at +1 732 885 6570,email: lili.lee@ge.com or Quality Assurance at +1 207 459 7557, email: heidi.kuell@ge.com.
- Initiated
- 2009-10-29
- Root cause
- Process control
- Terminated
- 2020-08-24
- Distribution
- Nationwide distribution: CA
- Recall number
- Z-2384-2010
- Recall status
- Terminated
- CDRH record ID
- 86584
- FDA classified
- 2010-09-07
- Recall event number
- 53819