Open original FDA record ↗Recall record
- FDA identifier
- Z-2384-2012
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Sep 14, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.
Source facts
- Action
- Philips Healthcare Imaging issued an Urgent - Field Safety Notice dated August 8, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The unit will be exchanged with a new unit that complies with the IEC requirements. Please reference Field Change Order (FCO) 71200081 when contacting I your local Philips representative. 1-800-722-9377. Customer were instructed to review the information contained in the letter with all members of their staff who need to be aware of the contents of the letter and to maintain a copy of the letter with the equipment instruction for Use. For further assistance, customers should contact their local Phillips representative: Technical Support Line 1-800-722-9377. For questions regarding this recall call 978-687-1501.
- Initiated
- 2012-08-14
- Root cause
- Device Design
- Terminated
- 2017-01-17
- Distribution
- Worldwide Distribution - USA including OH and WA and the country of Canada
- Recall number
- Z-2384-2012
- Recall status
- Terminated
- CDRH record ID
- 111935
- FDA classified
- 2012-09-14
- Recall event number
- 62884