Z-2384-2021: MY01 Continuous Compartmental Pressure Monitor, Reference Number MY01-0001

Recall record Z-2384-2021 for MY01, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2384-2021
Company in FDA record
MY01, INC.
FDA record creation date
Aug 27, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is an incorrect version of the needle in the introducer part of devices. The affected Lots were manufactured using needles with a needle trocar geometry which is out of specification and may require a slightly higher push force from the user to penetrate the patient skin.

Source facts

Action
A Medical Device Advisory Notice dated 6/28/21 was sent to customers. Clinicians Responsibilities/ Actions required: 1. Review this notification for awareness of the contents 2. Identify and quarantine any unused affected products in your inventory for return to MY01 Inc. Contact Customer Success team by phone (+1 (855) 292-6901) or email (help@my01.io) to arrange for your replacements. 3. There are no specific patient monitoring instructions related to this voluntary product removal that are recommended beyond your existing follow up schedule. 4. Complete Attachment 1- Certificate of acknowledgement and email to help@my01.io by 07-Jul-2021. 5. Include a copy of Attachment 1- Certificate of acknowledgement with product to be returned. Please use the return shipping label (to be provided by MY01 Inc.), please return affected product to: Regulatory Affairs and Quality Management MY01 Inc. 400 Maisonneuve Boulevard West, Suite 700, Montreal, Quebec, Canada H3A 1L4 6. Retain a copy of this document and Attachment 1- Certificate of acknowledgement with your records. 7. If you have further questions or concerns after reviewing this notice, please call the customer service team at 1-855-292-6901. Alternatively, your questions may be emailed to: help@my01.io Under 21 CFR 803, manufacturers are also required to report any serious injuries where a product has contributed or may have contributed to the event. Please keep MY01 Inc. informed of any adverse events associated with this by emailing help@my01.io or call +1(855)-292-6901.
Initiated
2021-06-28
Root cause
Under Investigation by firm
Terminated
2022-07-25
Distribution
US Nationwide distribution in the states of MD, OH, CA, NY.
Recall number
Z-2384-2021
Recall status
Terminated
CDRH record ID
188568
FDA record created
2021-08-27
Recall event number
88396

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗