There is incorrect barcode information, which can potentially cause the Alert Message "Calibrator Insert Missing" to be displayed.
Source facts
Action
The recalling firm issued an Urgent field Safety Notice, dated February 2010, to all affected customers informing them of the problem. Customers have been told to run calibrators from cups when calibrating or request a replacement lot. The notification should be forwarded to that this product was further distributed to. Siemens Technical Solutions Center should be contacted at 800-441-9250, if they are any technical questions.
Initiated
2010-02-26
Root cause
Employee error
Terminated
2010-10-13
Distribution
Worldwide Distribution -- USA, Austria, Germany, Netherlands, Italy, France, Denmark, Portugal, Canada, South Korea, Malaysia, Australia, and New Zealand.
Recall number
Z-2387-2010
Recall status
Terminated
CDRH record ID
89091
FDA classified
2010-09-09
Recall event number
54743