Open original FDA record ↗Recall record
- FDA identifier
- Z-2387-2011
- Company in FDA record
- Express Diagnostics Int'l., Inc.
- FDA classification date
- May 27, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Express Diagnostics Int'l, Inc. has initiated a medical device recall on various model numbers of DrugCheck devices, This recall was due to incorrect product expiration dating after findings from an FDA inspection. Use of expired product may not work properly, and may give incorrect screening results.
Source facts
- Action
- The firm, Express Diagnostic Int'l, Inc., sent an "Urgent Recall of DrugCheck Product" letter dated April 19, 2011 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to immediately quarantine the product; notify Express Diagnostic of the quantity even if no product is remaining; complete and return the DISTRIBUTOR/CUSTOMER RECALL CONFIRMATION FROM via fax to 1-507-526-2252 as soon as possible. In addition, if they have further distributed the product to identify their customers and notify them of the recall. Provided procedures to be followed for them to destroy the quarantined product. Express Diagnostics will replace the DrugCheck Drug Screen devices. If you have any questions, please contact the Quality Assurance Manager at 507-526-3951, ext. 117.
- Initiated
- 2011-03-18
- Root cause
- Incorrect or no expiration date
- Terminated
- 2012-07-06
- Distribution
- Worldwide distribution: USA (nationwide) including states of: FL, IA, MD, OK, and WA; and countries including: AUSTRALIA and ITALY.
- Recall number
- Z-2387-2011
- Recall status
- Terminated
- CDRH record ID
- 99781
- FDA classified
- 2011-05-27
- Recall event number
- 58652