Z-2388-2018: Discovery IGS 740 Interventional Fluoroscopic X-ray Systems

Recall record Z-2388-2018 for GE Medical Systems, SCS. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2388-2018
Company in FDA record
GE Medical Systems, SCS
FDA classification date
Oct 5, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The Discovery IGS and Innova IGS systems may experience X-ray abort errors which could result in the loss of imaging mode before or during fluoroscopic use.

Source facts

Action
GE Healthcare notified customers on approximately 03/29/2018, via an URGENT MEDICAL DEVICE CORRECTION letter. Safety instructions included that before each use, to ensure the IGS system is fully functional as indicated in the GE product labeling and follow each facility's established procedures before and during each use to manage the patient if a repeated X-ray abort error occurs during the procedure or in the case a complete loss of fluoroscopic viewing capability may occur. If the problem persists, customers are instructed to contact their local GE Healthcare Service Representative. Additionally, customers are instructed that GE Healthcare will correct all affected products at no cost and a GE Healthcare representative will contact each customer to arrange for the correction. Questions or concerns regarding the notification can be directed to GE Healthcare Service at 1-800-437-1171 or the local Service Representative.
Initiated
2018-03-29
Root cause
Process control
Terminated
2021-06-15
Distribution
US distribution: AR, CA, CA, FL, GA, IA, IL, NC, NH, NJ, NM, NY, PA, SC, TN, A,WI, WV. International Distribution: Argentina, Brazil, China, Colombia, Egypt, France, India, Indonesia, Iran, Italy, Japan, Kuwait, Lithuania, Myanmar, Palestine, Poland, Russia, Saudi Arabia, Singapore, Spain, Tunisia, Turkey, United Kingdom, Vietnam.
Recall number
Z-2388-2018
Recall status
Terminated
CDRH record ID
164965
FDA classified
2018-10-05
Recall event number
80177

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗