Skip to content

Public FDA record

Z-2388-2021: Jessie Sterile Arm Suspension Kit, Product Code 711300, (previous product code 100401), Sample Kit S71130

Recall record Z-2388-2021 for Aspen Surgical Products, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2388-2021
Company in FDA record
Aspen Surgical Products, Inc.
FDA record creation date
Aug 30, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Affected products were manufactured with materials found to have a bioburden level exceeding the limit at which the product's sterilization process was validated. As a result, the firm cannot provide full assurance that the affected products were effectively sterilized, and the product may compromise a sterile field and increase the risk of a surgical site infection.

Source facts

Action
Required Actions: 1. Check all storage and usage locations to confirm if you have any units of the affected item codes in your possession. Please note that these products bear an expiration date; any expired product should be discarded regardless of whether they are affected by this corrective action. 2. Distributors: Please check your inventory for affected products as indicated above. For products that you have distributed, please forward this notification to your customers who have received affected product. Please also provide to Aspen a list of customers who have received these products, including the number of each product and contact information for each customer, to patty.shook@aspensurgical.com. 3. Review, complete, sign and return the Acknowledgement Form which accompanies the notice.. 4. Destroy and discard any affected product found in your inventory. Your sales representative can assist with disposal of affected product if needed. 5. Contact Aspen Customer Service, at 888-364-7004 or customerservice@aspensurgical for information on credit options. At this time, we expect that replacement product will be available by October, 2021; however, we hope to expedite production and will contact you with updates on supply availability. In the interim, until Aspen suggests considering the following alternatives: o o Marco Shoulder Stabilization Kit: Smith & Nephew SPIDER Shoulder Stabilization Kit (7210573) o o Jessie Arm Suspension Kit: McConnell Arm Support (12-401) 6. Share this notice with others in your facility who need to be aware of this removal. Contact other facilities associated with your organization that may have received units of affected lots. 7. Maintain awareness of this notice until all affected product has been returned to Aspen Surgical or destroyed. 8. Keep a copy of this notice with affected product until returned to Aspen.
Initiated
2021-07-23
Root cause
Process control
Terminated
2022-07-26
Distribution
US nationwide distribution. No foreign distribution.
Recall number
Z-2388-2021
Recall status
Terminated
CDRH record ID
188753
FDA record created
2021-08-30
Recall event number
88476

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗