Open original FDA record ↗Recall record
- FDA identifier
- Z-2389-2008
- Company in FDA record
- Cardinal Health NeuroCare Division
- FDA classification date
- Sep 21, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Cardinal Health NeuroCare is implementing a voluntary field correction to replace a quantity of photic adapter cables, part number 085-463700, used to connect the NicLED Photic Strobe option, part number 698-625500, with NicoletOne EEG system. The photic adapter cable has a wiring error which delivers a reduced voltage and results in a decreased intensity of the NicLED photic. This reduced intensi
Source facts
- Action
- A "Cardinal Health Urgent: Medical Device Correction" letter was sent to consignees. The letter described the product and the problem. The consignees were asked to confirm completion of the cable exchange by returning an information card. If you have questions, contact Cardinal Health (formerly VIASYS) NeuroCare Technical Support Group at 1-800-356-0007 (US), 608-273-5000 (OUS) or email shane.sawall@cardinalhealth.com.
- Initiated
- 2008-06-05
- Root cause
- Nonconforming Material/Component
- Terminated
- 2010-10-13
- Distribution
- CA FL GA ID IN LA MD ME MI MO NC NY PA TN TX WA WI China Denmark Indonesia Japan Korea Mexico Netherlands Russia Saudi Arabia Slovenia
- Recall number
- Z-2389-2008
- Recall status
- Terminated
- CDRH record ID
- 72066
- FDA classified
- 2008-09-21
- Recall event number
- 48758