Skip to content

Public FDA record

Z-2390-2016: Stryker Orthopaedics Specialty Triathlon Tibial Alignment Handle with Secondary Lock, Non-Sterile, Used

Recall record Z-2390-2016 for Stryker Howmedica Osteonics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2390-2016
Company in FDA record
Stryker Howmedica Osteonics Corp.
FDA record creation date
Aug 11, 2016
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

It was reported that the secondary locking mechanism, and its corresponding components, disassociated from the Specialty Triathlon Tibial Alignment Handle with Secondary Lock Assembly during surgery. Upon further investigation, it was discovered that the weld, which was intended to hold the secondary locking mechanism together, did not meet the weld size specified on the engineering drawing.

Source facts

Action
Stryker sent an Urgent Product Recall letter dated June 24, 2016, with a Business Reply Form attached to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to inform users of the Urgent Product Recall letter and forward the notice to all those individuals who need to be aware within their organization. Customers were instructed to return all affected product at their location to: Stryker C/O Stericycle 2670 Executive Dr., Suite A Indianapolis, IN 46241 Customers were also instructed to complete and sign the enclosed Business Reply Form and fax it to 888-912-7352 or email to Stericycle at strykerortho4582@stericycle.com. Customers with questions were advised to call (201) 831-6693. For questions regarding this recall call 201-831-5000.
Initiated
2016-06-23
Root cause
Process control
Terminated
2018-08-01
Distribution
Nationwide Distribution to FL, IN, LA, NC, PA, TN and UT
Recall number
Z-2390-2016
Recall status
Terminated
CDRH record ID
148055
FDA record created
2016-08-11
Recall event number
74619

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗