Z-2390-2018: Elecsys Vitamin D total II, Vitamin D Test System, Material number 07464215160 Product Usage: This assay

Recall record Z-2390-2018 for Roche Diagnostics Corporation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2390-2018
Company in FDA record
Roche Diagnostics Corporation
FDA record creation date
Jul 8, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The device may give a falsely elevated result that is non-reproducible. If vitamin D level is monitored by laboratory tests and adapted based on the result, a dosage reduction of vitamin D supplementation cannot be excluded, and as a consequence, vitamin D deficiency may develop over time.

Source facts

Action
On April 27, 2018, the firm issued an Urgent Medical Device Correction letter to all affected customers. The letter advised customers in the following workaround to improve sample quality prior to measurement with the Elecsys Vitamin D total II assay. The firm stated these steps are to be carried out temporarily until further information is provided: 1. Perform a thorough inspection of your pre-analytical handling to ensure a correct assay performance of Elecsys Vitamin D total II and to comply with the individual specifications of the primary tube manufacturers for all tubes in use (centrifugation conditions are particularly important as well as the elimination of foam/bubbles from the surface of the sample). 2. If the issue persists, switch from plasma to serum. 3. If you still encounter the issue for plasma or serum, re-centrifuge the sample in a secondary tube for 10 minutes at 2000 x g prior to measurement with the Elecsys Vitamin D total II assay. If you are still experiencing issues after completing the steps above, discontinue the use of the Elecsys Vitamin D Total II assay. Contact the Roche Support Network Customer Support Center at 1-800-428-2336. For the MODULAR ANALYTICS E 170 module, cobas e 411 analyzer, and cobas e 601 and 602 analyzers, the availability of the Elecsys Vitamin D Assay (catalog number 06506780160) has been extended and may be used as an alternate method until this issue is resolved. Customers were asked to do the following: " Follow the workaround outlined in this Urgent Medical Device Correction (UMDC) to improve sample quality prior to measurement with the Elecsys Vitamin D total II assay until further information is provided. " If none of the workaround actions listed on the previous page successfully resolve the issue, please discontinue use of the Elecsys Vitamin D total II assay and contact the Roche Support Network Customer Support Center at 1-800-428-2336. " For all analyzers except the cobas e 801 module, the V
Initiated
2018-04-27
Root cause
Under Investigation by firm
Terminated
2020-05-11
Distribution
US Nationwide distribution.
Recall number
Z-2390-2018
Recall status
Terminated
CDRH record ID
164204
FDA record created
2018-07-08
Recall event number
80008

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗