Open original FDA record ↗Recall record
- FDA identifier
- Z-2391-2020
- Company in FDA record
- Medtronic Sofamor Danek USA Inc
- FDA record creation date
- Jun 16, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Product may not maintain its setting characteristics for the labeled shelf-life duration. Use may lead to longer time for product to set at the time of implantation which may increase the required surgical and anesthesia time for the patient. If the device seeps or creeps from its intended position prior to setting and creates incomplete void or gap filling, possible longer-term health consequences could include incomplete bone formation or lack of bone formation, delayed union, or nonunion.
Source facts
- Action
- Medtronic notified customers on about 04/24/2020 via "URGENT: Voluntary Market Removal (Recall)" letter. Customers were informed that Medtronic was notified by Skeletal Kinetics LLC that product that may not maintain its setting characteristics for the labeled shelf-life duration of the product, which may lead to a longer period of time required for the product to set at the time of implantation. Instructions included to not use the product and locate and removed the impacted product from inventory, complete the provided Customer Confirmation Form and return it via email to rs.sdmriskmgt@medtronic.com, and contact your Medtronic Sales Representative to coordinate the return and replacement of affected product.
- Initiated
- 2020-04-24
- Root cause
- Under Investigation by firm
- Terminated
- 2021-03-31
- Distribution
- Nationwide distribution to CA, FL, IL, IN, KY, MN, MO, MT, OH, PA, TX, VA, WA.
- Recall number
- Z-2391-2020
- Recall status
- Terminated
- CDRH record ID
- 181342
- FDA record created
- 2020-06-16
- Recall event number
- 85611