Open original FDA record ↗Recall record
- FDA identifier
- Z-2392-2016
- Company in FDA record
- Hitachi Medical Systems America Inc
- FDA record creation date
- Aug 11, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The firm discovered that the centrifugal force applied to the power supply may cause it to fail over time. The exam performed at the time of the failure will likely be lost, resulting in the need to rescan the patient.
Source facts
- Action
- Hitachi sent an Device Correction Notice dated April 1, 2015, to all affected customers. The letter identified the product the problem and the action to be taken by the customer. Customers were informed that there is a problem with the Hitachi Medical Systems America, Inc.CT system; the issue is with a power supply in the CT gantry sub-system. During fast rotation (i.e., 35 sec), the centrifugal force applied to the power supply may cause it to fail over time. For further questions, please call (330) 425-1313 ext. 3720.
- Initiated
- 2015-06-30
- Root cause
- Component design/selection
- Terminated
- 2016-11-01
- Distribution
- US Distribution to the states of : CA, FL, GA, IA, IL, IN, MD, NC, NY, OH, PA, SD, TN, TX & WY., and Internationally to Brazil and Mexico.
- Recall number
- Z-2392-2016
- Recall status
- Terminated
- CDRH record ID
- 148276
- FDA record created
- 2016-08-11
- Recall event number
- 74714