Skip to content

Public FDA record

Z-2393-2020: DDU-2400EN Lifeline PRO, UDI 00815098020270 - Product Usage: Automated External Defibrillators (AED) are

Recall record Z-2393-2020 for Defibtech, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2393-2020
Company in FDA record
Defibtech, LLC
FDA record creation date
Jun 17, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A component issue may cause the AED to abort a shock delivery or reset unexpectedly, which may cause an affected device to fail to delivery shock, and/or to fail to deliver shock in range.

Source facts

Action
The firm contacted affected customers via email and mailed letter on April 27, 2020. Distributors were asked to forward letters to end users. Customers were informed that affected products may abort a shock delivery or reset unexpectedly. Customers were asked to return all affected products for screening. Actions to be taken by the Customer/User: 1. If you have additional AEDs on hand , please quarantine and return any affected units as identified in the Attachment for screening or replacement. Do not include the pads and battery packs when you return your unit. 2. If no other AEDs are available, please continue using the device with caution until a loaner or replacement is provided. 3. Ensure that all users of the AED are informed of the product issue. Defibtech will provide support for handling all aspects of the return of identified units for screening. The firm will contact customers to help them with the process. If you have questions now, please contact the firm using the email address FA2020@defibtech.com. Additionally, Defibtech Customer Support is available by calling 1-877-453-4507, 8:30 A.M. to 5:30 P.M. (Eastern), Monday - Friday.
Initiated
2020-04-27
Root cause
Component design/selection
Terminated
2021-01-08
Distribution
Worldwide distribution - US Nationwide distribution and the countries of Australia, Canada, France, Malaysia, Mauritius, Netherlands, Thailand, United Arab Emirates.
Recall number
Z-2393-2020
Recall status
Terminated
CDRH record ID
181571
FDA record created
2020-06-17
Recall event number
85702

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗