Open original FDA record ↗Recall record
- FDA identifier
- Z-2393-2023
- Company in FDA record
- Cytrellis Biosystems, Inc.
- FDA classification date
- Aug 11, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to prolonged healing and irregularities in patient skin texture post-treatment, and/or scarring
Source facts
- Action
- Cytrellis Biosystems, Inc issued Urgent Medical Device Recall letter via email on 6/23/23 and revised letter on 7/7/23. Letter states reason for recall, health risk and action to take: A member of our service engineering team will be reaching out to your practice to schedule a visit to conduct the software update to your console on site. The visit will take approximately 30-45 minutes. Our goal for the completion of updates for all existing customers is by August 31st. Actions to be taken by the Customer: 1) If you drop or have dropped your handpiece, stop immediate use of the handpiece and contact Cytrellis. 2) Complete the attached Recall Return Response Form and provide it to the service engineering representative that conducts the software update on the device located at your site as acknowledgment of this recall. Any questions can be directed to medicalinformation@cytrellis.com. Website: www.cytrellis.com
- Initiated
- 2023-06-23
- Root cause
- Process control
- Distribution
- Nationwide
- Recall number
- Z-2393-2023
- Recall status
- Open, Classified
- CDRH record ID
- 202174
- FDA classified
- 2023-08-11
- Recall event number
- 92689