Z-2393-2023: ellacor System with Micro-Coring Technology-Indicated for use by medical professionals for the treatment

Recall record Z-2393-2023 for Cytrellis Biosystems, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-2393-2023
Company in FDA record
Cytrellis Biosystems, Inc.
FDA classification date
Aug 11, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to prolonged healing and irregularities in patient skin texture post-treatment, and/or scarring

Source facts

Action
Cytrellis Biosystems, Inc issued Urgent Medical Device Recall letter via email on 6/23/23 and revised letter on 7/7/23. Letter states reason for recall, health risk and action to take: A member of our service engineering team will be reaching out to your practice to schedule a visit to conduct the software update to your console on site. The visit will take approximately 30-45 minutes. Our goal for the completion of updates for all existing customers is by August 31st. Actions to be taken by the Customer: 1) If you drop or have dropped your handpiece, stop immediate use of the handpiece and contact Cytrellis. 2) Complete the attached Recall Return Response Form and provide it to the service engineering representative that conducts the software update on the device located at your site as acknowledgment of this recall. Any questions can be directed to medicalinformation@cytrellis.com. Website: www.cytrellis.com
Initiated
2023-06-23
Root cause
Process control
Distribution
Nationwide
Recall number
Z-2393-2023
Recall status
Open, Classified
CDRH record ID
202174
FDA classified
2023-08-11
Recall event number
92689

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗